Analytical HPLC / UHPLC

KNAUER AZURA® HPLC 862 bar system

KNAUER AZURA® HPLC 862 bar system

key features

- With the 862 bar system, you choose the most widespread system solution of the KNAUER AZURA family.

- We recommend this system for all classical applications and column types, as well as for faster and optimized methods with shorter columns and run times.

- The complete solution for your daily HPLC routine. Easily upgradable to an ultra low dispersion chromatography (ULDC) system

- Covers most analytical applications due to optimal continuous working conditions in the range of 0.1–4.0 ml/min at up to 862 bar

- Use classical HPLC columns just as well as fast analysis columns packed with 3 µm particles or even core shell materials

- Typical applications include food, pharmaceutical and environmental analyzes

 

Specifications

Back pressure range - 862 bar max. at up to 2 ml/min, 400 bar at up to 10 ml/min

Flow rate range - 0.001–10 ml/min

Recommended continuous flow rate - Up to 5 ml/min

Available pump systems - Isocratic/Binary HPG/Quaternary LPG

Biocompatibility - Optionally available for backpressure range up to 400 bar

Injection options - Autosampler, Liquid handler, manual, automatted via electrical injection valve

Injection volume range - 0.1 µl–5000 µl depending on sample loop

Sample capacity - 108 standard HPLC vials (1.5 ml) or up to 768 positions when equipped with microtiter plates, with liquid handler up to 390 HPLC vials

Column thermostatting range - 5–85 °C

Eluent preheating - Optional cartridge for eluent pre-heating

Column switching options - Optional switching between up to 8 columns

Choice of detectors - DAD 190–700 nm, DAD 190–1000 nm, RID, UVD, MWD, FLD, ELSD, MS

Choice of software packages - ClarityChrom® CDS, OpenLAB® CDS EZChrom Edition, Mobile Control, Chromeleon™ 7.2

Request quote