System key features
The high-speed solution for your daily HPLC routine. Ideally suited for high throughput applications.
We recommend this system for UHPLC applications whenever your column runs at back pressures above 800 bar. This system is suitable for all analytical column types, but designed especially for high-speed methods with short columns and run times.
- The complete solution for your daily UHPLC routine
- Covers ultra high pressure applications due to optimal continuous working conditions in the range of 0.1–2.0 ml/min at up to 1000 bar
- Use classical HPLC columns just as well as high resolution UHPLC columns packed with 2 µm particles or core shell materials
- Typical applications include food, pharmaceutical and environmental analyses
Specifications
Back pressure range - 1000 bar max.
Flow rate range - 0.001–5 ml/min
Recommended continuous flow rate - Up to 2 ml/min
Available pump systems - Binary HPG/Quaternary LPG
Injection options - Autosampler, Liquid handler, manual, via electrical injection valve
Injection volume range - 0.1 µl–5000 µl depending on sample loop
Sample capacity - 108 standard HPLC vials (1.5 ml) or up to 768 positions when equipped with microtiter plates, with liquid handler up to 390 HPLC vials
Column thermostatting range- 5–85 °C
Eluent preheating - Optional cartridge for eluent pre-heating
Column switching options - Optional switching between up to 8 columns
Choice of detectors - DAD 190–700 nm, DAD 190–1000 nm, RID, UVD, MWD, FLD, ELSD, MS
Choice of software packages - ClarityChrom® CDS, OpenLAB® CDS EZChrom Edition, Mobile Control, Chromeleon™ 7.2